4.3 Setting the Test Frequency
______________________________________________________________________________________________
4.3.2 Selecting the Input Method
Changes to the Ten key screen
Ten Key Screen
Digit Screen
Changes to the Digit screen
Test frequency
(Hz)
Setting range
Test range
Test voltage (V, CV)
Test current (CC)
42.0 - 100.0 k
0.1
Ω
- 100 M
Ω
0.010 V- 5.000 V
0.01 mA - 99.99 mA
100.1 k - 1.000 M
0.1
Ω
- 1 M
Ω
1.001 M - 5.000 M
0.1
Ω
- 100 k
Ω
0.010 V - 1.000 V
0.01 mA - 20.00 mA
1.
On the Initial screen, press the
key, and the Menu screen will be
displayed.
2.
Press the
key on the Menu screen, and the display will change over to
the one of the Frequency setting screens which was previously used.
3.
The display can be switched over between these two Frequency setting
screens just by pressing the soft keys shown in the figure.
Some test ranges and test signal levels are unavailable at certain test
frequencies. These ranges and levels can be checked using the table below.
If the test frequency is set to a specific value, the maximum values are
automatically selected when the test range and test signal level settings
exceed the values specified in the table below. For example, when the test
frequency is set to 1.001 MHz in a test range of 1 M
Ω
and a test signal
level of 1.5 V, the test range is automatically changed to 100 k
Ω
and the
test signal level to 1 V.