![Hioki 3532-50 Instruction Manual Download Page 151](http://html1.mh-extra.com/html/hioki/3532-50/3532-50_instruction-manual_2137204151.webp)
137
_____________________________________________________________________________________________
5.6 Testing Using EXT I/O
______________________________________________________________________________________________
5.6.1 Normal Testing
EOM
T5
T4
TRIG
T1
INDEX
T2
T3
HI IN LO AND
Decision
result
Previous decision result (Note 1)
LD1
LD2
LD3
LD4
LD5
Current test conditions
Panel number 9
(Panel load signals)
Following test conditions
Panel number 22
T6
(Comparator
result output)
(Measurement
finished signal)
(Analog measurement
in progress signal)
(Testing start signal)
T7
NOTE
5.6 Testing Using EXT I/O
With the test conditions for testing by the comparator having been set (with
the trigger setting set to external trigger), when in this state a trigger signal
is input via the EXT I/O connector, or when on the screen the
soft
key is pressed, then the decision result is output on the comparator result
output signal line of the EXT I/O connector.
When a panel number has been selected using the panel load signal at the
input of a trigger signal via the EXT I/O connector, the 3532-50 loads the
measurement conditions for the panel number and then begins measurement.
An example of testing timing is as follows:
You can use a communication command (:IO:RESult:RESet) to select whether
the judgment results for comparator measurement are reset when the start-of-
measurement signal is input or updated when measurement ends.
Note1: Reset at the same time as TRIG
―――――
: HIGH
Not reset at the same time as TRIG
―――――
: Last judgment result remains
(Initial setting)