4.3 Setting the Test Frequency
______________________________________________________________________________________________
4.3.3 Input Using the Ten Key Screen
Shows the test
frequency input so far
Returns to the Initial screen
Units keys (shown in non-reversed video, until
the numerical value for the frequency has been input)
Numeric keypad
Shows the currently
set test frequency
NOTE
1.
Access to the following Ten key screen.
2.
Input the new test frequency directly using the numeric keypad.
Frequency range which can be set: 42.0 Hz to 5.000 MHz
If you attempt to set a test frequency greater than 5 MHz, it will
automatically be reduced to 5 MHz.
If you attempt to set a test frequency lower than 42 Hz, it will automatically
be increased to 42 Hz.
If you make a mistake during input, press the
key to cancel the input
so far, and start again.
3.
When you have completed inputting the numerical value for the frequency,
press the appropriate units key to confirm the displayed frequency.
If you change screens before confirming the frequency you have set, the
frequency will not be changed.
Until the numerical value for the frequency has been input, the units keys
cannot be used.
4.
When the setting has been established as above, press the
key to return
to the Initial screen.