40
_____________________________________________________________________________________________
4.4 Setting the Test Signal Level
______________________________________________________________________________________________
4.4.2 Selecting the Level Type
CAUTION
Be absolutely sure not to change over between constant current
(CC) level and constant voltage (CV) level with the test
terminals still connected to the sample to be tested. Doing so
may damage the test sample.
Initial screen
Menu screen
When the setting has
been established as
above, press this key
to return to the Initial
screen.
The digit keys (If held down continuously, the corresponding digit increments
or decrements continuously.) (If the limit for the corresponding digit is reached,
these keys change to non-reversed video and become inactive.)
The key corresponding to the currently established type of
test signal level is shown in non-reversed video.
Monitor display portion
Test frequency
Test signal level
Voltage between test
samples
Limit value
Current flowing
through the test
sample
NOTE
1.
Press the
key on the Initial screen to display the Menu screen.
2.
Press the
key on the Menu screen to display the Test signal level
setting screen.
The current situation can be checked using the monitor display.
3.
The method of control for the test signal level can be selected from the
following three:
Set open circuit voltage (V) The value of the open circuit voltage is set.
Set constant voltage (CV) The value of the voltage between the
terminals of the object under test is set.
Set constant current (CC) The value of the current flowing through the
object under test is set.
The accuracy of testing varies according to the test signal level. Check the
accuracy table in Section 7.5.