With certain faults, the unit will switch off all functions. Take the
unit out of service and submit it to a Service Engineer for
inspection and to re-establish the operational reliability.
There is a risk of the patient may be overheated or
undercooled.
► Monitor the patient's body temperature when using the unit
and the water pad on the patient.
Do not start the unit if
► the display has failed,
► individual segments of the temperature display have failed
(temperature can no longer be read safely),
► the yellow fault lamp (13) lights permanently or not at all
(functional test),
► the signal tone continues to sound or does not sound at all
(functional test),
► the unit does not respond to key operation,
► the unit does not respond as described in chapter 6.2.3
"Functional test" when switching on or when running a
function test.