Guidelines and declaration of manufacturer
HICO-VARIOTHERM 550
68
Table 2
Guidance and manufacturer declaration – Electromagnetic immunity
The HICO-Variotherm 550 is designed for operation in the electromagnetic environment listed below.
The customer or user of the HICO-Variotherm 550 should ensure that the device is used in such an environment.
Emission tests
IEC 60601 test level
Compliance level
Electromagnetic
environment - guidance
Electrostatic discharge
(ESD) according to IEC
61000-4-2
±
6 kV contact discharge
±
8 kV air discharge
±
6 kV contact discharge
±
8 kV air discharge
Floors should be wood,
concrete or ceramic tile. If the
floor is covered with synthetic
material, the relative humidity
should be at least 30 %.
Fast transient electrical
interferences variables
/ bursts
according to IEC
61000-4-4
±
2 kV for mains cables
±
1 kV for input and
output cables
±
2 kV for mains cables
±
1 kV for input and
output cables
The quality of the power
supply should correspond to
that of a typical commercial or
hospital environment.
Surge voltage/Surges
according to
IEC 61000-4-5
±
1 kV differential mode
voltage
±
2 kV common mode
voltage
±
1kV differential mode
voltage
±
2 kV common mode
voltage
The quality of the power
supply should correspond to
that of a typical commercial or
hospital environment.
Voltage dips, short
interruptions and
variations of the power
supply
according to IEC
61000-4-11
<5 % U
T
(>95 % dip in U
T
)
for ½ cycles
40 % U
T
(60 % dip in U
T
)
for 5 cycles
70 % U
T
(60 % dip in U
T
)
for 25 cycles
<5 % UT
(>95 % dip in U
T
)
for 5 s
<5 % U
T
(>95 % dip in U
T
)
for ½ cycles
40 % U
T
(60 % dip in U
T
)
for 5 cycles
70 % U
T
(60 % dip in U
T
)
for 25 cycles
<5 % U
T
(>95 % dip in U
T
)
for 5 s
The quality of the power
supply should correspond to
that of a typical commercial or
hospital environment. If the
user of the HICO-Variotherm
550 demands a continued
operation during interruptions
of the power supply, we
recommend to supply the
HICO-Variotherm 550 from an
uninterruptible power supply or
a battery.
Power frequency
(50/60 Hz) Magnetic
field according to
IEC 61000-4-8
3 A/m
3 A/m
Magnetic fields at the mains
frequency should correspond
to that of a typical commercial
and hospital environment
Note: UT is the alternating power supply voltage prior to application of the test level