![Hico Medical Systems HICO-VARIOTHERM 550 Instructions For Use Manual Download Page 54](http://html1.mh-extra.com/html/hico-medical-systems/hico-variotherm-550/hico-variotherm-550_instructions-for-use-manual_2129330054.webp)
Cleaning and disinfecting
HICO-VARIOTHERM 550
54
7.3 Water pads, hose extension
If possible use decalcified water. Only wipe the surfaces with a
damp cloth. For cleaning use only warm water (max. 50 °C) with
a mild commercial dish washer liquid. Wipe off again with clear
water and wipe the surface dry with a cloth.
To disinfect the surface we recommend an area, wiping or
spraying disinfection according to VAH-list
(e.g. with S&M
Mikrozid AF). When using a disinfectant follow the instructions of
the manufacturer.
Only use the pad and the hose extension again after the
disinfectant has evaporated completely.
Check the pad and the hose extension for damage, deformation
or cracks; replace damaged accessories.
Do not apply any other cleaning or decontamination processes
other than the ones recommended by Hirtz & Co. KG.
Before introducing new methods, check these together with Hirtz
& Co. KG.
Only this makes sure that the unit will not be damaged by this
method.
17
V
erband für
A
ngewandte
H
ygiene, list available at mhp-Verlag, Wiesbaden