Brief instruction
HICO-VARIOTHERM 550
71
13 Brief instruction
•
Connect the unit to the mains supply
•
Couple the water pad to the unit.
•
Check the water level on the unit.
•
Switch on the unit by the mains switch and watch the
automatic function test.
•
If the nominal temperature value is >38 °C or <35 °C when
switching on the unit, an alarm will be emitted. Acknowledge
and start by pressing the release key once.
•
Set the temperature with the arrow keys.
•
For temperatures higher than 38 °C, press the arrow key and
release key together.
•
For temperatures below 35 °C, press arrow key and release
key together.
•
Depending on the application, place the water pad under or
over the patient.
•
Monitor the body temperature of the patient.
•
Monitor the water level and water flow on the unit.
•
In case of continuous operation perform a manual function test
once every day by pressing the "Function test" key.
•
Acoustic warnings of medium priority (“WATER LEVEL!?” and
“TEMP.DIFF >1 °C”) can be interrupted using the "Audio
paused" key.
•
All other acoustic alarms of medium priority cannot be
interrupted. Switch off the unit at the mains (16).