Commissioning
HICO-VARIOTHERM 550
19
4.3 Connecting the HICO-VARIOTHERM 550
Danger caused by water in connection with electricity.
► Only connect the unit to the mains supply after it has been
filled.
4.3.1 Filling the system
•
Unscrew the cap from the water filling socket (5), e.g. with a
coin. Take care not to lose the sealing ring of the cap.
•
Observe the water level indicator (6) while filling. After filling,
the water level should be slightly below the MAX mark.
•
Fill unit tank with sterile filtered tap water
, to which a
germicidal agent
•
The usage of a non-proportional share of disinfectants can
reduce the lifetime of those parts of the unit that will come into
contact with water!
For initial filling of the unit, please follow the
Danger caused by water in connection with electricity.
Water is electrically conductive.
► If water has been spilled when filling the unit, the unit must
first be thoroughly dried and should only be connected to the
mains supply and switched on after it has completely dried
out.
2
Depending on the quality of the water (e.g. calcareousness among other things), the lifetime of the unit parts that
come into contact with water is reduced.
7
e.g. 10 ml SANOSIL mixture (from Sanosil, Farchant)
(1000 ml sterile filtered tap water + 10 ml Sanosil
solution). Alternatively, another long-term germinating preservative can also be used, e.g. Micropur or Certisil. For
further information, please contact Hirtz & Co.KG.
The use of a disproportionately high amount of disinfectant can reduce the lifetime of unit parts that come into
contact with water!