Guidelines and declaration of manufacturer
HICO-VARIOTHERM 550
67
12 Guidelines and declaration of manufacturer
Guidelines and declaration of manufacturer – electromagnetic emissions
Tables for medical electrical equipment general Information:
Table 1
Guidance and manufacturer declaration – Electromagnetic emission
The HICO-Variotherm 550 is designed for operation in the electromagnetic environment
listed below. The customer or user of the HICO-Variotherm 550 should ensure that the
device be used in such an environment.
Emission test
Agreement
Electromagnetic environment - guidance
RF emissions according to
CISPR 11
Group 1
The HICO-Variotherm 550 uses RF energy
exclusively for its internal function.
Therefore, its RF emission is very small. It is
unlikely that nearby electronic devices
will be disturbed.
RF emissions according to
CISPR 11
Class B
The HICO-Variotherm 550 is designed for use in all
premises, including such
residential areas that are directly connected to a
public power supply that also
supplies residential buildings.
Harmonics according to
IEC 61000-3-2
Class A
Voltage fluctuations/
flicker according to IEC 61000-3-3
correspond