Commissioning
HICO-VARIOTHERM 550
18
4.2.3 Installation with five-castor stand (optional)
The five-castor stand available as an accessory makes the
HICO-VARIOTHERM 550, designed as a stationary unit, mobile.
Install the unit together with the stand as follows:
•
Assemble the five-castor stand as described in its operating
instructions.
•
Place the stand plate (1) on the vertical pipe (2) of the five-
castor stand.
•
Fix the plate with the star screw (3).
•
There is a safety screw (4) in the stand plate; place the HICO-
VARIOTHERM 550 on the plate in such a way that the screw
and the thread in the base of the unit are aligned with one
another.
•
Secure the unit by screwing the safety screw into the base of
the unit.
Personal and material damage may occur when installing
and using the five-castor stand with the unit!
► Five-castor stand assembly: For the safe functioning and
stable condition of the system, it must be ensured that, when
assembling the five-castor stand, the holding rod (2) in the
centre section and all 5 stabilisers with castors are safely
mounted in accordance with the instructions and safety
information on the five-castor stand.
► The five-castor stand only serves to carry HICO systems!
► Do not use any aggressive detergents; the five-castor stand
is not suitable for use in washing systems!
The five-castor stand should be checked for full functionality of all
components at least once a year:
► Are the screw connections still secure?
► Cleanse the castors of contaminants so that the brakes work
and the electrical conductivity is not marred.
► This should be checked directly after assembling and at
regular intervals during the operating period (at least with / at
the recommended yearly maintenance interval)
► If a fault is detected, then only original spare parts may be
used and only replaced by an expert or customer service.