Cleaning and disinfecting
HICO-VARIOTHERM 550
51
7.2 Unit
7.2.1 Water circulation
7.2.1.1.
General Information for cleaning the water circuit
The cleaning process consists of a total of three steps, all of
which have to be performed in a regular cycle of 14 days. First
the basic cleaning of the unit and the application parts takes
place. This is followed by disinfection, afterwards flushing and
refilling. You have to wear disposable gloves during the entire
cleaning process and avoid contact with the sterile filtered tap
and the cleaning agent.
7.2.1.2.
Basic cleaning
Disconnect the unit from the power supply and remove any
accessories that may be connected. Empty the unit and all
accessories (e.g. water pads, extension hoses, etc.) entirely.
Refill the unit with a neutral cleaner (observe manufacturer’s
instructions) and use only sterile filtered tap water for possible
dilution. Connect the parts to be cleaned to the unit and connect
the unit to the power supply. Use the setpoint buttons to set the
temperature to 35 °C. Turn on the unit and leave it on for 30
minutes. Disconnect the unit from the power supply and remove
the connected accessories. Empty the unit and all accessories
entirely.
7.2.1.3.
Disinfection
For disinfection, any VAH
-listed disinfectant which is free of
phenolic derivatives can be used. Suitable are e.g. Terralin
protect, Gigasept FF, Gigasept AF, Mikrozid AF or Sanosil S003.
Observe the dosage recommendation of the respective
manufacturer of the disinfectant. If necessary, prepare a
disinfecting solution according to the instructions of the
manufacturer of the respective disinfectant. Fill the unit with the
disinfectant or the disinfecting solution. Connect the device to the
power supply and connect it to any disinfected application parts.
Turn on the unit and allow it to run for about two minutes so that
the disinfectant can spread throughout the circuit; if necessary,
refill the disinfectant or disinfecting solution. Switch off the unit and
observe the predefined exposure time of the manufacturer of the
disinfectant. Afterwards disconnect the unit from the power supply
and empty the unit and all accessories entirely.