Control and operation
HICO-VARIOTHERM 550
34
6.2.3 Function test
After switching on the device performs a function test. During this
test observe whether all indicating elements behave as described
below:
•
a short audible alarm indicates that the unit is prepared for a
possible power failure alarm
The unit now checks its self-sufficient protection and
•
shows the result in the display (12)
•
the temperature displays (11) and (14) shows
•
the fault lamp (13) lights up
•
the audible alarm is on.
This test takes a few seconds.
If the functional tests did not complete successfully, the unit
is no longer fail safe.
Do not operate unit if
► the unit does not issue the short signal tone for the power
failure alarm when switching on.
► the automatic functional test automatically switches the unit
off because it has detected a defect in the independent
protection device.
► one or more displays are defective.
In these cases, have the unit checked by HICO customer
service.