![Hico Medical Systems HICO-VARIOTHERM 550 Instructions For Use Manual Download Page 17](http://html1.mh-extra.com/html/hico-medical-systems/hico-variotherm-550/hico-variotherm-550_instructions-for-use-manual_2129330017.webp)
Commissioning
HICO-VARIOTHERM 550
17
4.2.2 Setting up with the unit carrier (optional)
The stand, which is optionally available as an accessory, makes
the stationary HICO VARIOTHERM 550 mobile.
Install the unit together with the stand as follows:
•
Assemble the stand as described in its operating instructions
(K2B0045).
•
There are two elongated holes in the base of the stand; place
the HICO-VARIOTHERM 550 on the base of the stand in such
a way that the elongated holes and the thread in the base of
the unit are aligned with one another.
•
The base of the stand is screwed to the unit using a knurled
screw.
Personal and material damage may occur when installing
and using the stand with the unit!
► To assemble the stand, observe the safety notes in the
assembly and operating instructions on the stand!
► The stand only serves to carry HICO systems!
► Do not use any aggressive detergents; the stand is not
suitable for use in washing systems!
► The stand should be checked for full functionality of all
components at least once a year:
Are the screw connections still secure?
Cleanse the castors of contaminants so that the brakes work
and the electrical conductivity is not marred.
If a fault is detected, only original spare parts may be used and
only replaced by an expert or customer service.