43
When the battery is too low, the oximeter gives a medium alarm
for low battery.
After the low battery alarm occurs, the oximeter can still be
operated for a few minutes before it turns off automatically.
The low battery icon
displays on LCD, and gives a
medium alarm. (Make sure
Alarm System
in menu is ON.)
5.2.5 Higher than Hi Alarm Limit
If the measured SpO
2
or PR value is higher than the Hi Alarm
(upper alarm limit), the oximeter gives a high alarm.
Here we take PR for example:
If the measured PR value is higher than the setup
PR Hi Alarm
,
the oximeter gives a high alarm (Make sure
Alarm System
in
menu is
ON
). A
↑
icon displays near PR, which indicates that the
measured value is higher than that of
PR Hi Alarm
, it will
synchronously flash with PR value.
5.2.6 Lower than Lo Alarm Limit
If the measured SpO
2
or PR value is lower than the
Lo Alarm
(lower alarm limit), the oximeter gives a high alarm.
Here we take SpO
2
for example:
If the measured SpO
2
value is lower than the set
SpO
2
Lo Alarm
,
the oximeter gives a low SpO
2
alarm. (Make sure
Alarm System
in menu is
ON
.)
A
↓
icon displays near SpO
2
value, which indicates the
measured value is lower than that of
SpO
2
Lo Alarm
, it will
synchronously flash with SpO
2
value.
5.2.7 Alarm Silence
If
Alarm System
in menu is
ON
, pressing the
alarm silence
button, the audio alarm will be off for the pause period set by the
user, but the visual alarm is still active.
When the audio alarm is off, press the
Alarm Silence
button to
reactivate the audio alarm function.
5.2.8 Turning off Alarm System
Summary of Contents for HS-Vet
Page 1: ......
Page 65: ...64 Appendix Ⅲ Record Table ID No Name Time SpO2 PR NOTE ...
Page 98: ...97 Configuración del sistema Configuración de alarma ...
Page 136: ...135 pulsioxímetro ...
Page 139: ...138 Apéndice III Tabla de registro Nº ID Nombre Hora SpO2 PR NOTA ...
Page 204: ...203 Annexe Ⅲ Record Table No ID Nom Temps SpO2 PR REMARQUE ...
Page 237: ...236 Alarm Setup ...
Page 278: ...277 Anhang III Tabelle für Aufzeichnungen ID Nr Bezeichnung Zeit SpO2 PF HINWEIS ...