A2.2 Electromagnetic Immunity - for all
EQUIPMENT and SYSTEMS
Guidance and manufacturer’s declaration – electromagnetic
immunity
The oximeter is intended for use in the electromagnetic
environment specified below. The customer or the user of the
oximeter should assure that it is used in such an environment.
Emissions test
Compliance Complianc
e level
Electromagneti
c environment -
guidance
Electrostatic
discharge(ESD)
IEC/EN61000-4-2
±6kV contact
±8kV air
±6kV
contact
±8kV air
Floors should be
wood, concrete
or ceramic tile. If
floors are
covered with
synthetic
material, the
relative humidity
should be at least
30%.
Electrical Fast
Transient/Burst
IEC/EN61000-4-4
±2kV for
power supply
lines
±1kV for
input/output
lines (>3m)
N/A
Mains power
quality
should be that of
a typical
commercial or
hospital
environment.
Surge
IEC/EN61000-4-5
line to line
line to ground
N/A