6 Performance Considerations
6.1 Performance Verification
Qualified service personnel are responsible for performance
verification procedures before the oximeter is used for the first
time in a clinical setting.
6.2 Oximeter Performance Considerations
There are some patient conditions that can affect the oximeter’s
measurements.
Dysfunctional Hemoglobins
Dysfunctional hemoglobins, such as carboxyhemoglobin,
methemoglobin, and sulfhemoglobin, are unable to carry
oxygen. SpO
2
readings may appear normal; however, a
patient may be hypoxic because less hemoglobin is
available to carry oxygen. Further assessment beyond
pulse oximeter is recommended.
Anemia
Anemia causes decreased arterial oxygen content.
Although SpO
2
readings may appear normal, an anemic
patient may be hypoxic. Correcting anemia can improve
arterial oxygen content. The oximeter may fail to provide
SpO
2
if hemoglobin levels fall below 5 gm/dl.
Saturation
The oximeter displays saturation level between 1% and
100%.
Pulse rate
The oximeter displays pulse rate between 30 and 254
beats per minute (bpm). The sensor accuracy ranges do
not apply to pulse rates above 254 bpm.
Data update period
The data update period is two seconds typically, and 10
seconds in extreme conditions.
6.3 Sensor Performance Considerations
Summary of Contents for HS-Vet
Page 1: ......
Page 65: ...64 Appendix Ⅲ Record Table ID No Name Time SpO2 PR NOTE ...
Page 98: ...97 Configuración del sistema Configuración de alarma ...
Page 136: ...135 pulsioxímetro ...
Page 139: ...138 Apéndice III Tabla de registro Nº ID Nombre Hora SpO2 PR NOTA ...
Page 204: ...203 Annexe Ⅲ Record Table No ID Nom Temps SpO2 PR REMARQUE ...
Page 237: ...236 Alarm Setup ...
Page 278: ...277 Anhang III Tabelle für Aufzeichnungen ID Nr Bezeichnung Zeit SpO2 PF HINWEIS ...