36
the upper limit of PR Alarm is from “1 + the lower
limit of PR Alarm” to 350.
Press the
Up Arrow
or
Down Arrow
button to
increase or decrease values.
Choose
PR Lo Alarm
in menu, press the
Function
button enter setup. The
PR Lo Alarm
box changes
from real line to broken line. The adjustable range for
the lower limit of PR Alarm is from 0 to “the upper
limit of PR Alarm – 1”.
Press the
Function
button, confirm the alarm range
setup.
Hi Alarm
is always higher than
Lo Alarm
by at least
1 bpm.
Press the
On/Off
button to exit the menu, and return to
measurement state.
3.3.11 Patient ID No. Setup
The oximeter can support 100 IDs, and 300-hour data storage.
When entering menu, press the
Function
button to set ID (valid
range is from 1 to 100). The ID display box on the interface will
change from real line to broken line.
After choosing ID, press the
Function
button to confirm the setup.
If the ID exists, the following confirmation dialog box will pop
up.
Figure 3-13 Confirm to Cover Data
3.3.12 Data Storage
Choose
Data Storage
, and set it to
ON
, then the measurement
data can be stored.
During the data storage, ID can not be changed. If the user wants
Summary of Contents for HS-Vet
Page 1: ......
Page 65: ...64 Appendix Ⅲ Record Table ID No Name Time SpO2 PR NOTE ...
Page 98: ...97 Configuración del sistema Configuración de alarma ...
Page 136: ...135 pulsioxímetro ...
Page 139: ...138 Apéndice III Tabla de registro Nº ID Nombre Hora SpO2 PR NOTA ...
Page 204: ...203 Annexe Ⅲ Record Table No ID Nom Temps SpO2 PR REMARQUE ...
Page 237: ...236 Alarm Setup ...
Page 278: ...277 Anhang III Tabelle für Aufzeichnungen ID Nr Bezeichnung Zeit SpO2 PF HINWEIS ...