3.3.8 Alarm System
Set
Alarm System
to
ON
or
OFF
to turn on or off the alarm
system.
If
Alarm system
is set to
OFF
, a dialog box pops up as follows:
Figure 3-12 Confirm to Turn off Alarm
If
Alarm System
is
ON
and an alarm occurs, the oximeter will
give a visual alarm and an audio alarm.
Pressing the
Alarm Silence
button can suspend the alarm system
for seconds (the pause period can be set to 30, 60, 90 or 120s by
the user, see section 3.3.5), the audio alarm off icon displays. But
the visual alarm is still active. For example, if the measured SpO
2
value is higher than
SpO
2
Hi Alarm
or lower than
SpO
2
Lo
Alarm
, there will be
↑
or
↓
icon displayed on screen, and the SpO
2
or PR character will flash.
If
Alarm system
is set to
OFF
, all audio alarms and visual alarms
are turned off.
WARNING
When the Alarm system is off, the oximeter will not give an
alarm prompt. In order to avoid endangering the patient’s life,
the user should use this function cautiously.
3.3.9 SpO2 Alarm Setup
The user can choose
SpO
2
Hi Alarm
and
SpO
2
Lo Alarm
in
menu to adjust SpO
2
alarm limit. Press the
Up Arrow
button or
Down Arrow
button to increase or decrease alarm limit.
By default,
SpO
2
Hi Alarm
and
SpO
2
Lo Alarm
are
100
and
85
respectively.
Summary of Contents for HS-Vet
Page 1: ......
Page 65: ...64 Appendix Ⅲ Record Table ID No Name Time SpO2 PR NOTE ...
Page 98: ...97 Configuración del sistema Configuración de alarma ...
Page 136: ...135 pulsioxímetro ...
Page 139: ...138 Apéndice III Tabla de registro Nº ID Nombre Hora SpO2 PR NOTA ...
Page 204: ...203 Annexe Ⅲ Record Table No ID Nom Temps SpO2 PR REMARQUE ...
Page 237: ...236 Alarm Setup ...
Page 278: ...277 Anhang III Tabelle für Aufzeichnungen ID Nr Bezeichnung Zeit SpO2 PF HINWEIS ...