8 Appendix
8.2 Interface XS2 Reservoir
Only the tanks and product supply devices provided by LOCTITE can be operated at this
interface!
Pin no.:
Function
Addition
1
"Refill" signal input
Not galvanically isolated
2
"Empty" signal input
3
4
5
Internal ground output
6
Tank valve output
7
8
Tank detection input
9
Internal 24V
Pins 5 and 9 are used to supply the tank electronics.
The internal power supply is rated for a maximum of 100 mA. Under no circumstances may a
power supply be connected to these outputs!
RC15 DC Drive Controller
Reservoir 97106
Reservoir 97108
1
2
3
4
5
6
7
8
9
1
2
3
4
5
6
7
8
9
Reservoir refill
Reservoir empty
0VDC
Reservoir
solenoid valve
Reservoir
identification signal
+24VDC output
Ground
+24VDC,
50 mA
Connect original Loctite product supply only!