184
Symptom
Explanation
Action
No transfer of
data
1) No network communication
with OR or PC.
2) Communication error.
3) No communication.
4) No communication between
the first and second Docking
Station.
5) No communication between
the Secondary Docking
Stations.
6) No communication with any
Docking Station.
1a) Check that the LED on the
Docking Station does not show a
flashing red light.
1b) Check the network configuration
of PDS/PDS
+
.
1c) If multiple Docking Stations are
used, make sure not to mix
Docking Stations labelled PDS
+
/
SDS
+
with unlabelled Docking
Stations.
1d) Check the external set-up and
settings (OR/PC, firewall, network
infrastructure).
2a) More than 5 Docking Stations are
docked together.
2b) If multiple Docking Stations are
used, make sure not to mix
Docking Stations labelled PDS
+
/
SDS
+
with unlabelled Docking
Stations.
3a) Check that a USB connection is
not used for communication in a
Primary Docking Station.
3b) If multiple Docking Stations are
used, make sure not to mix
Docking Stations labelled PDS
+
/
SDS
+
with unlabelled Docking
Stations.
4) Check that two Primary Stations
are not docked together.
5a) Check the electrical connections.
The LED on the docking station
should give either a stable green
light or a flashing green light.
5b) If multiple Docking Stations are
used, make sure not to mix
Docking Stations labelled PDS
+
/
SDS
+
with unlabelled Docking
Stations.
6) Check the electrical connection.
The LED on the Docking Station
should give either a stable green
light or a flashing green light.
If problems remain contact
distributor.