127
6.1.2 QC Lockout
If the required QC Test is not performed
within the time or number of measurements
indicated by the QC Reminder,
the Analyzer will perform a QC lockout.
FIGURE 6-3
The QC Lockout (1) indicates that the
Analyzer has performed a QC lockout.
Patient Tests are not allowed when the
Analyzer has performed a QC lockout.
Linearity and Proficiency Tests are not
allowed when the Analyzer has performed a
QC lockout.
STAT Tests are not allowed when the
Analyzer has performed a QC lockout if
usage of STAT Tests is set to “Not Used” (see
) or if the number
of STAT Tests performed has reached the
max. limit allowed.
The Information text (2) indicates the types
of QC Tests required to unlock the QC
Lockout.
a) To enter the QC Test procedure, press the
QC Test button (3).
1 QC Lockout
2 Information text
3 QC Test button
08/09/03
10:55 AM
JAMES SMITH
Run QC 1, QC 2
1
3
2
Summary of Contents for 201 DM
Page 1: ...HemoCue 201 DM Analyzer HemoCue 201 DM Docking Station Reference Manual ...
Page 6: ...PART I Introduction ...
Page 41: ...PART II Settings ...
Page 108: ...PART III Routine use ...
Page 181: ...PART IV Other actions ...
Page 192: ...PART V Miscellaneous ...