179
Symptom
Explanation
Action
E27
The QC memory is full. No more
QC data can be saved.
Save test results by placing the
Analyzer in a Docking Station
connected to DMS Software or OR.
For the following, supervisory
authority is needed.
a) Delete all or part of the QC Tests
stored in the Analyzer (see
)
b) Change Analyzer settings
regarding full internal memory for
QC tests to “Overwrite” (see
E28
The Analyzer log memory is full.
No more Error Codes and Log
Notes can be saved.
Save test results by placing the
Analyzer in a Docking Station
connected to DMS Software or an
OR.
For the following, supervisory
authority is needed.
a) Delete all or part of the Analyzer
logs stored in the Analyzer (see
b) As the Analyzer settings regarding
full memory for Analyzer logs is
the same as the setting for full
memory of QC tests, change
Analyzer behaviour for QC tests
to “Overwrite” (see
).
E29
The electronic self-test failed.
The communication self-test
failed. The Analyzer may not
work properly when connected
to a docking station.This is stored
as a failed Electronic QC Test
(EQC) in the Analyzer Log book.
Analyzer needs service. Contact
distributor.