133
FIGURE 6-10
The Numeric Test Result (5) and the
Qualitative Test Result (6) are displayed in
bold text.
For a result within the Approved area (2) (the
blank area), the Qualitative Test Result (6) will
indicate “Pass”.
For a result within the Warning area (3) (the
dotted area), the Qualitative Test Result (6)
will indicate “Pass, Warning”.
For a result within the Fail area (4) (the solid
area), or for two consecutive results within
the Warning area (3), the Qualitative Test
Result will (6) indicate “Fail”.
To avoid or unlock a QC lockout, the
Qualitative Test Result must indicate “Pass”.
e) To view a graphic presentation of the
most recent QC Tests, press the Statistics
button (7).
will be displayed.
f) Press the Confirm button (8).
or
will be displayed.
2 Approved area
3 Warning area
4 Fail area
5 Numeric Test Result
6 Qualitative Test Result
7 Statistics button
8 Confirm button
QC Level 1
Lot No: 234567
Result
50 mg/dL
Pass
43 47 53 57
6
2
3
5
4
7
8