44
Unit:mg/dL
Warning Min (0) 30
Warning Max (500) 60
QC Level 1
1
5
2
8
7
4
6
9
3
0
11
55
22
88
77
44
66
99
33
00
10
11
12
FIGURE 3-11
The warning limits define the range of
results for QC testing that must be obtained
to prevent a warning from being displayed.
j) Enter min. value (10) and max. value (11)
for the warning interval.
k) Press the Confirm button (12).
A similar image for the lockout limits will be
displayed. The lockout limits define the
range for QC testing that must be obtained
to prevent the Analyzer from performing a
lockout.
l) In the image for the lockout limits, enter
min. value (10) and max. value (11) for the
lockout interval.
m) Press the Confirm button (12).
10 Min. value
11 Max. value
12 Confirm button