![HBK DSE Operating Manual Download Page 17](http://html1.mh-extra.com/html/hbk/dse/dse_operating-manual_2111691017.webp)
15
DSE
INSTALLATION
X1, sensor connection
The DSE has an 8-pin M12 plug (A-coded) for connecting the sensor. This is located on
the front of the DSE, and is labeled X1 ([1] in Fig. 5.1). The pin assignment is shown in
Fig. 5.2.
X1: Sensor connection, A-coded plug
1
Measurement
2
NC (Not connected)
3
Sense lead +
4
NC
5
Sense lead –
6
Bridge excitation
voltage –
7
Bridge excitation
v
8
Measurement signal
–
1
2
3
4
5
6
7
8
Fig. 5.2
X1, sensor connection, front view of plug
Important
You can connect strain gage sensors in a 6-wire configuration with a resistance between
250 and 4500 Ω. If you connect a load cell in a 4-wire configuration, it is essential to
connect “Sense lead +” to “Bridge excitation v” and “Sense lead -” to “Bridge
excitation voltage -”, such as by means of jumpers on the sensor or in the plug (Fig. 5.4).
Connection for strain gage full bridge in 6-wire configuration
3
2
1
4
rd
wh
gy
bk
gn
bl
Cable wire colors (HBM transducer):
bl = blue; gn = green; gy = gray; rd = red; bk = black; wh = white
Measurement
Sense lead
*
Bridge excitation v
Sense lead +
Bridge excitation voltage
*
Cable shields on
connector housing
Measurement signal
*
1
2
3
4
5
6
7
8
Fig. 5.3
6-wire connection
Connect the cable shields extensively to the connector housing.
Summary of Contents for DSE
Page 1: ...DSE ENGLISH DEUTSCH Operating Manual Bedienungsanleitung ...
Page 3: ...DSE ENGLISH DEUTSCH Operating Manual ...
Page 22: ...DSE INSTALLATION 20 Fig 5 7 DSE in star topology left and ring topology right ...
Page 41: ...39 DSE CONNECTINGDSE VIA TCP IP Fig 6 24 DSE signals ...
Page 102: ...DSE ENGLISH DEUTSCH Bedienungsanleitung ...
Page 122: ...21 DSE INSTALLATION Abb 5 7 DSE in Sterntopologie links und Ringtopologie rechts ...
Page 203: ...DSE ENTSORGUNG 102 ...
Page 204: ...HBK Hottinger Brüel Kjaer www hbkworld com info hbkworld com A05529 01 X00 02 ...