16
Connection for strain gage full bridge in 4-wire configuration
3
2
1
4
rd
wh
bk
bl
Measurement
Bridge excitation v
Bridge excitation voltage
*
Measurement signal
*
1
2
3
4
5
6
7
8
Feedback bridges
Cable wire colors (HBM transducer):
rd = red; wh= white; bk= black; bu= blue
Cable shields on
connector housing
Fig. 5.4
4-wire connection
For 4-wire transducers, you need to solder feedback bridges (shorting bars) for the sense
leads into the transducer plug. If you need an extension, make it in a 6-wire configura
tion. Connect the cable shields extensively to the connector housing.
X3, power supply connection
The DSE has an 4-pin M12 plug (T-coded) for connecting the power supply. This is
located in the middle on the back, and is labeled X3 ([4] in Fig. 5.1). Note that the
operating voltage may be between 15 V
DC
and a maximum of 30 V
DC
. The nominal
(rated) voltage is 24 V
DC
. The pin assignment is shown in Fig. 5.5.
X3: Power supply, T-coded plug
1
Supply v
2
NC
3
GND
4
NC
1
2
4
3
Fig. 5.5
X3, supply voltage connection, front view of plug
Summary of Contents for DSE
Page 1: ...DSE ENGLISH DEUTSCH Operating Manual Bedienungsanleitung ...
Page 3: ...DSE ENGLISH DEUTSCH Operating Manual ...
Page 22: ...DSE INSTALLATION 20 Fig 5 7 DSE in star topology left and ring topology right ...
Page 41: ...39 DSE CONNECTINGDSE VIA TCP IP Fig 6 24 DSE signals ...
Page 102: ...DSE ENGLISH DEUTSCH Bedienungsanleitung ...
Page 122: ...21 DSE INSTALLATION Abb 5 7 DSE in Sterntopologie links und Ringtopologie rechts ...
Page 203: ...DSE ENTSORGUNG 102 ...
Page 204: ...HBK Hottinger Brüel Kjaer www hbkworld com info hbkworld com A05529 01 X00 02 ...