![HBK DSE Operating Manual Download Page 80](http://html1.mh-extra.com/html/hbk/dse/dse_operating-manual_2111691080.webp)
DSE
INDUSTRIAL ETHERNET
78
Explanation
No.
Byte
Format
R/W
Type
Module
Check
weigher
extended
Mean value
check
weigher
RO
F32
4
Mean value of weighing results
Standard
deviation
check
weigher
RO
F32
4
Standard deviation of weighing
results
Minimum
value check
weigher
RO
F32
4
Minimum weighing result
Maximum
value check
weigher
RO
F32
4
Maximum weighing result
Result sam
ple count
check
weigher
RO
U16
2
Number of weighing results
Objects 0x6002 and 0x6020 have the following attributes in detail:
Command word “Command 6002”, object 0x6002/01.
Command
[String]
Command
[U32]
Function
idle
0x656c6469
None, except to set the reply to 'idle'.
calz
0x7a6c6163
Sets the zero point of the empty scale (scale
dead load), and so starts the adjustment
procedure. Must be followed by caln or exit.
caln
0x6e6c6163
Sets the maximum capacity and ends the
adjustment procedure.
exit
0x74697865
Terminates the adjustment procedure
prematurely.
tare
0x65726174
Tare. Sets the net value to zero.
zero
0x6f72657a
Zero balancing. Sets the gross value to zero.
Reply to “Command 6002”, object 0x6002/02
Summary of Contents for DSE
Page 1: ...DSE ENGLISH DEUTSCH Operating Manual Bedienungsanleitung ...
Page 3: ...DSE ENGLISH DEUTSCH Operating Manual ...
Page 22: ...DSE INSTALLATION 20 Fig 5 7 DSE in star topology left and ring topology right ...
Page 41: ...39 DSE CONNECTINGDSE VIA TCP IP Fig 6 24 DSE signals ...
Page 102: ...DSE ENGLISH DEUTSCH Bedienungsanleitung ...
Page 122: ...21 DSE INSTALLATION Abb 5 7 DSE in Sterntopologie links und Ringtopologie rechts ...
Page 203: ...DSE ENTSORGUNG 102 ...
Page 204: ...HBK Hottinger Brüel Kjaer www hbkworld com info hbkworld com A05529 01 X00 02 ...