![HBK DSE Operating Manual Download Page 54](http://html1.mh-extra.com/html/hbk/dse/dse_operating-manual_2111691054.webp)
DSE
WEB INTERFACE
52
General settings
Fig. 7.9
Amplifier menu
The following settings are available:
S
Decimal Point: Select the desired number of decimal places. The maximum number of
digits before and after the decimal place that can be displayed is 8. If your selection
exceeds this value, it will be discarded.
S
Resolution: Select the desired resolution for your measuring range. The selection
options are adjusted depending on the number of decimal places. In multi-range and
multi-interval applications, the resolution selected here is used for the first measuring
range. The second - and where appropriate third - stages are displayed in the next
highest resolution.
S
Weight motion detection: Enter the desired value for standstill detection (Weight
motion detection) here. The number is displayed in d per second. The quantity d is
Summary of Contents for DSE