Firmware Version: 01.020 and later
English
S C P I P r o g r a m m e r s M a n u a l
H M S S e r i e s
1
HAMEG Instruments GmbH • Industriestraße 6 • D-63533 Mainhausen
Subject to change without notice
Tel.: +49 (0) 6182 800 0 • Fax: +49 (0) 6182 800 100 • E-Mail: [email protected]
Geschäftsführer: Dipl.-Ing. Holger Asmussen • AG Offenbach am Main HRB 41200
Firmware Version: 01.020 and later
English
S C P I P r o g r a m m e r s M a n u a l
H M S S e r i e s
HMO352x, HMO2524
HMO72x ... HMO202x
Firmware Version: 03.00 and later
English
:ACQuire
:CHAN1OFFSet
:DISPlay
:HCOPy
*ESR?
*OPC?
:LOGic:SIZE
:MEASure
:POD1
:SYSTem
*SRE?
S C P I P r o g r a m m e r s M a n u a l
H M O S e r i e s
SCPI Commands HMC8012
DATA:LOG:MODE {UNLimited | COUNt | TIME}
Selects the data logging mode.
Parameters:
<Mode>
UNLimited
Infinite data capture
COUNt
Number of measurement values to be captured
TIME
Duration of the measurement values capture
*RST:
UNL
DATA:LOG:MODE?
Returns the data logging mode.
Return values:
<Mode>
UNL
Infinite data capture
COUN
Number of measurement values to be captured
TIME
Duration of the measurement values capture
DATA:LOG:TIME
<
time in seconds
>
Sets the data logging time.
Parameters:
<Time>
Internal:
50000h
max.
External:
Defined by USB stick capacity
DATA:LOG:TIME?
Returns the duration of the measurement values capture.
Return values:
<Time>
e.g. 5.04
DATA:LOG:COUNt
<
no of samples
>
Sets the number of measurement values to be captured.
Parameters:
<Samples>
Internal:
50000
max.
External:
Defined by USB stick capacity
DATA:LOG:COUNt?
Returns the number of measurement values to be captured.
Return values:
<Time>
e.g. 1.00E+01
DATA:LOG:INTerval
<
interval in seconds
>
Selects a logging measurement interval. The measurement interval describes the time between the recorded
measurements.
Parameters:
<Interval>
3600s max.
*RST: 0s