34
HAMEG Instruments GmbH • Industriestr. 6 • D-63533 Mainhausen • Deutschland
Subject to change without notice
Tel.: +49 (0) 6182 800 0 • Fax: +49 (0) 6182 800 100 • E-Mail: [email protected]
Geschäftsführer: Dipl.-Ing. Holger Asmussen, Dipl.-Ing. Andre Vander Stichelen • AG Offenbach am Main HRB 41200
1
1
HAMEG Instruments GmbH • Industriestraße 6 • D-63533 Mainhausen
Subject to change without notice
Tel.: +49 (0) 6182 800 0 • Fax: +49 (0) 6182 800 100 • E-Mail: [email protected]
Geschäftsführer: Dipl.-Ing. Holger Asmussen • AG Offenbach am Main HRB 41200
Firmware Version: 01.020 and later
English
S C P I P r o g r a m m e r s M a n u a l
H M S S e r i e s
1
HAMEG Instruments GmbH • Industriestraße 6 • D-63533 Mainhausen
Subject to change without notice
Tel.: +49 (0) 6182 800 0 • Fax: +49 (0) 6182 800 100 • E-Mail: [email protected]
Geschäftsführer: Dipl.-Ing. Holger Asmussen • AG Offenbach am Main HRB 41200
Firmware Version: 01.020 and later
English
S C P I P r o g r a m m e r s M a n u a l
H M S S e r i e s
HMO352x, HMO2524
HMO72x ... HMO202x
Firmware Version: 03.00 and later
English
:ACQuire
:CHAN1OFFSet
:DISPlay
:HCOPy
*ESR?
*OPC?
:LOGic:SIZE
:MEASure
:POD1
:SYSTem
*SRE?
S C P I P r o g r a m m e r s M a n u a l
H M O S e r i e s
SCPI Commands HMC8012
[SENSe:]CAPacitance:NULL:VALue {
<
Value
>
|MIN|MAX}
Sets the null value of the capacitance measurement function.
Parameters:
<Value>
0 to 500µF (adjustable in 1fF steps)
MIN: 0F
MAX: 500µF
*RST: 0F
[SENSe:]CAPacitance:NULL:VALue? [{MIN|MAX}]
Returns the null value of the capacitance measurement function.
Return values:
<Value>
0 to 500µF
MIN: 0.0E+00
MAX: 5.000000E-04
[SENSe:]CAPacitance:RANGe:AUTO
<
Mode
>
Activates or disables the auto range mode of the capacitance measurement function.
Parameters:
<Mode>
ON | OFF | 1 | 0
ON:
Activates the auto range mode of the capacitance measurement function
OFF:
Disables the auto range mode of the capacitance measurement function
1:
Activates the auto range mode of the capacitance measurement function
0:
Disables the auto range mode of the capacitance measurement function
*RST: ON
[SENSe:]CAPacitance:RANGe:AUTO?
Returns the auto mode state of the capacitance measurement function.
Return values:
<State>
1 | 0
1:
Auto range of the capacitance measurement function is activated
0:
Auto range of the capacitance measurement function is disabled
[SENSe:]CAPacitance:RANGe[:UPPer] {
<
Range
>
|MIN|MAX|DEF}
Sets the capacitance measurement function range.
Parameters:
<Range>
5nF, 50nF, 500nF, 5µF, 50µF, 500µF
MIN: 5nF
MAX: 500µF
DEFault: 5nF
*RST: AUTO