55
HAMEG Instruments GmbH • Industriestr. 6 • D-63533 Mainhausen • Deutschland
Subject to change without notice
Tel.: +49 (0) 6182 800 0 • Fax: +49 (0) 6182 800 100 • E-Mail: [email protected]
Geschäftsführer: Dipl.-Ing. Holger Asmussen, Dipl.-Ing. Andre Vander Stichelen • AG Offenbach am Main HRB 41200
1
1
HAMEG Instruments GmbH • Industriestraße 6 • D-63533 Mainhausen
Subject to change without notice
Tel.: +49 (0) 6182 800 0 • Fax: +49 (0) 6182 800 100 • E-Mail: [email protected]
Geschäftsführer: Dipl.-Ing. Holger Asmussen • AG Offenbach am Main HRB 41200
Firmware Version: 01.020 and later
English
S C P I P r o g r a m m e r s M a n u a l
H M S S e r i e s
1
HAMEG Instruments GmbH • Industriestraße 6 • D-63533 Mainhausen
Subject to change without notice
Tel.: +49 (0) 6182 800 0 • Fax: +49 (0) 6182 800 100 • E-Mail: [email protected]
Geschäftsführer: Dipl.-Ing. Holger Asmussen • AG Offenbach am Main HRB 41200
Firmware Version: 01.020 and later
English
S C P I P r o g r a m m e r s M a n u a l
H M S S e r i e s
HMO352x, HMO2524
HMO72x ... HMO202x
Firmware Version: 03.00 and later
English
:ACQuire
:CHAN1OFFSet
:DISPlay
:HCOPy
*ESR?
*OPC?
:LOGic:SIZE
:MEASure
:POD1
:SYSTem
*SRE?
S C P I P r o g r a m m e r s M a n u a l
H M O S e r i e s
SCPI Commands HMC8012
[SENSe:]RESistance:RANGe[:UPPer] {
<
Range
>
|MIN|MAX|DEF}
Sets the 2-wire (2w) resistance measurement function range.
Parameters:
<Range>
400
Ω
, 4k
Ω
, 40k
Ω
, 400k
Ω
, 4M
Ω,
40M
Ω,
250M
Ω
MIN: 400
Ω
MAX: 250M
Ω
DEFault: 400
Ω
*RST: AUTO
[SENSe:]RESistance:RANGe[:UPPer]? [{MIN|MAX}]
Returns the state of the 2-wire (2w) resistance measurement function range.
Return values:
<Range>
400
Ω
, 4k
Ω
, 40k
Ω
, 400k
Ω
, 4M
Ω,
40M
Ω,
250M
Ω
MIN: 4.0002
MAX: 2.5008
Temperature configuration commands
CONFigure:TEMPerature [{<Probe_Type>|DEF}[,{<Type>|DEF}[,1] ........................... 55
[SENSe:]TEMPerature:NULL[:STATe] {ON|OFF} ......................................................... 56
[SENSe:]TEMPerature:NULL[:STATe]? ....................................................................... 56
[SENSe:]TEMPerature:NULL:VALue {<Value>|MIN|MAX} ......................................... 56
[SENSe:]TEMPerature:NULL:VALue? [{MIN|MAX}] ................................................... 56
[SENSe:]TEMPerature:TRANsducer:RTD:TYPE <Type> ............................................ 56
[SENSe:]TEMPerature:TRANsducer:RTD:TYPE? ....................................................... 57
[SENSe:]TEMPerature:TRANsducer:TYPE <Probe_Type> ........................................ 57
[SENSe:]TEMPerature:TRANsducer:TYPE? ............................................................... 57
UNIT:TEMPerature {C | K | F} ...................................................................................... 57
UNIT:TEMPerature? ................................................................................................... 57
CONFigure:TEMPerature [{
<
Probe_Type
>
|DEF}[,{
<
Type
>
|DEF}[,1]
Configures the instrument for 2-wire / 4-wire temperature measurements, but does not initiate a measurement. Use
the READ? command to query the 2-wire / 4-wire temperature measurement value with specified range or the CONF?
command to query the 2-wire resistance function configuration.
Parameter:
<Probe_Type>
FRTD | RTD
FRTD: 4-wire temperature value (4W)
RTD:
2-wire temperature value (2W)
DEF:
RTD (2W)
<Type>
PT100 | PT500 | PT1000
DEF:
PT100
Example:
CONF:TEMP FRTD,PT500