60
HAMEG Instruments GmbH • Industriestr. 6 • D-63533 Mainhausen • Deutschland
Subject to change without notice
Tel.: +49 (0) 6182 800 0 • Fax: +49 (0) 6182 800 100 • E-Mail: [email protected]
Geschäftsführer: Dipl.-Ing. Holger Asmussen, Dipl.-Ing. Andre Vander Stichelen • AG Offenbach am Main HRB 41200
1
1
HAMEG Instruments GmbH • Industriestraße 6 • D-63533 Mainhausen
Subject to change without notice
Tel.: +49 (0) 6182 800 0 • Fax: +49 (0) 6182 800 100 • E-Mail: [email protected]
Geschäftsführer: Dipl.-Ing. Holger Asmussen • AG Offenbach am Main HRB 41200
Firmware Version: 01.020 and later
English
S C P I P r o g r a m m e r s M a n u a l
H M S S e r i e s
1
HAMEG Instruments GmbH • Industriestraße 6 • D-63533 Mainhausen
Subject to change without notice
Tel.: +49 (0) 6182 800 0 • Fax: +49 (0) 6182 800 100 • E-Mail: [email protected]
Geschäftsführer: Dipl.-Ing. Holger Asmussen • AG Offenbach am Main HRB 41200
Firmware Version: 01.020 and later
English
S C P I P r o g r a m m e r s M a n u a l
H M S S e r i e s
HMO352x, HMO2524
HMO72x ... HMO202x
Firmware Version: 03.00 and later
English
:ACQuire
:CHAN1OFFSet
:DISPlay
:HCOPy
*ESR?
*OPC?
:LOGic:SIZE
:MEASure
:POD1
:SYSTem
*SRE?
S C P I P r o g r a m m e r s M a n u a l
H M O S e r i e s
SCPI Commands HMC8012
Measurement
function
Mathematic functions
Stats
(AVER)
Limits
NULL
dB
dBm
Power
DC V
●
●
●
●
●
●
(only with 2nd function DC I)
AC V
●
●
●
●
●
−
DC I
●
●
●
●
●
●
(only with 2nd function DC I)
AC I
●
●
●
●
●
−
Ω
●
●
●
−
−
−
CAP
●
●
●
−
−
−
SENSOR
●
−
●
−
−
−
−
−
−
−
−
−
Tab. 2.1: Measurement functions with possible math options
CALCulate:FUNCtion?
Returns the calculation function.
Return values:
<Function>
NULL:
Null function
DB:
dB function (available in DC V AC V, DC I, AC I)
DBM:
dBm function (available in DC V AC V, DC I, AC I)
AVER:
Statistic measurements
LIM:
Limit lines
POW:
DC power value (available in DC V/DC I or DC I/DC V mode)
CALCulate[:STATe] {OFF | ON}
Turns with the CALC FUNC command selected calculation function on or off.
Parameters:
<State>
ON:
The selected calculation function is activated.
OFF:
The selected calculation function is disabled.
CALCulate[:STATe]?
Returns the state of the CALC FUNC command selected calculation function.
Return values:
<State>
1 | 0
1
ON - Calculation function is activated.
0
OFF - Calculation function is disabled.
CALCulate:POWer?
Returns the DC power value. To activate the power display function use the CALC:FUNC POW resp. the CALC ON
command.
Return values:
<Value>