5
HAMEG Instruments GmbH • Industriestr. 6 • D-63533 Mainhausen • Deutschland
Subject to change without notice
Tel.: +49 (0) 6182 800 0 • Fax: +49 (0) 6182 800 100 • E-Mail: [email protected]
Geschäftsführer: Dipl.-Ing. Holger Asmussen, Dipl.-Ing. Andre Vander Stichelen • AG Offenbach am Main HRB 41200
1
1
HAMEG Instruments GmbH • Industriestraße 6 • D-63533 Mainhausen
Subject to change without notice
Tel.: +49 (0) 6182 800 0 • Fax: +49 (0) 6182 800 100 • E-Mail: [email protected]
Geschäftsführer: Dipl.-Ing. Holger Asmussen • AG Offenbach am Main HRB 41200
Firmware Version: 01.020 and later
English
S C P I P r o g r a m m e r s M a n u a l
H M S S e r i e s
1
HAMEG Instruments GmbH • Industriestraße 6 • D-63533 Mainhausen
Subject to change without notice
Tel.: +49 (0) 6182 800 0 • Fax: +49 (0) 6182 800 100 • E-Mail: [email protected]
Geschäftsführer: Dipl.-Ing. Holger Asmussen • AG Offenbach am Main HRB 41200
Firmware Version: 01.020 and later
English
S C P I P r o g r a m m e r s M a n u a l
H M S S e r i e s
HMO352x, HMO2524
HMO72x ... HMO202x
Firmware Version: 03.00 and later
English
:ACQuire
:CHAN1OFFSet
:DISPlay
:HCOPy
*ESR?
*OPC?
:LOGic:SIZE
:MEASure
:POD1
:SYSTem
*SRE?
S C P I P r o g r a m m e r s M a n u a l
H M O S e r i e s
SCPI Commands HMC8012
By default, the instrument is configured to use dynamic TCP/IP configuration and obtain all address information au-
tomatically. This means that it is safe to establish a physical connection to the LAN without any previous instrument
configuration.
Risk of network errors
Connection errors can affect the entire network. If your network does not support DHCP, or if you choose
to disable dynamic TCP/IP configuration, you must assign valid address information before connecting the
instrument to the LAN. Contact your network administrator to obtain a valid IP address.
Configuring LAN parameters
1. Press the
SETUP
key and then the
Interface
softkey.
2. Press the
Ethernet
and then the
Parameter
softkey.
Note: By default, the instrument is set to not use DHCP. If the instrument is set to use DHCP and cannot find a DHCP
server, it takes about two minutes until the
Ethernet
menu is available.
The „Ethernet Settings“ dialog box is displayed.
Fig. 1.1: Ethernet Settings dialog box
Some data is displayed for information only and cannot be edited. This includes the „MAC“ (physical) address of the
connector and the „Link“ status information.
4. Define the IP address of the instrument by entering each of the four blocks individually (manual mode) or choose the
automatic IP-Mode.
a) In manual mode (MAN) define the first block number using the knob.
b) Press
Next
to move to the next block and define the number.
c) When the IP address is complete, press
Down
to continue with the next setting.
5. Define the „Subnetmask“ and „Gateway“ in the same way.
6. Select the „RAW Port“ - the port number for SCPI socket communication.
7. Select the „VXI-11- Port“ used by the instrument.
8. Select the „Transfer“ mode. This mode can either be determined automatically („Auto“ setting), or you can select a
combination of a transfer rate and half or full duplex manually.
9. Press
Save
to save the LAN parameters.