Firmware Version: 01.020 and later
English
S C P I P r o g r a m m e r s M a n u a l
H M S S e r i e s
1
HAMEG Instruments GmbH • Industriestraße 6 • D-63533 Mainhausen
Subject to change without notice
Tel.: +49 (0) 6182 800 0 • Fax: +49 (0) 6182 800 100 • E-Mail: [email protected]
Geschäftsführer: Dipl.-Ing. Holger Asmussen • AG Offenbach am Main HRB 41200
Firmware Version: 01.020 and later
English
S C P I P r o g r a m m e r s M a n u a l
H M S S e r i e s
HMO352x, HMO2524
HMO72x ... HMO202x
Firmware Version: 03.00 and later
English
:ACQuire
:CHAN1OFFSet
:DISPlay
:HCOPy
*ESR?
*OPC?
:LOGic:SIZE
:MEASure
:POD1
:SYSTem
*SRE?
S C P I P r o g r a m m e r s M a n u a l
H M O S e r i e s
SCPI Commands HMC8012
[SENSe:]CAPacitance:RANGe[:UPPer]? [{MIN|MAX|DEF}]
Returns the state of the capacitance measurement function range.
Return values:
<Range>
5nF, 50nF, 500nF, 5µF, 50µF, 500µF
MIN: 5.00000000E-09
MAX: 5.00000000E-04
DEF: 5.00000000E-09
Continuity configuration commands
CONFigure:CONTinuity .............................................................................................. 35
[SENSe:]CONTinuity:THReshold {<Threshold>|MIN|MAX|DEF} ................................. 35
[SENSe:]CONTinuity:THReshold? [{MIN|MAX|DEF}] .................................................. 35
[SENSe:]CONTinuity:BEEPer[:STATe] {ON|OFF} ......................................................... 36
[SENSe:]CONTinuity:BEEPer[:STATe]? ....................................................................... 36
CONFigure:CONTinuity
Configures the instrument for continuity measurements, but does not initiate a measurement. Use the READ? com-
mand to query the continuity measurement value or the CONF? command to query the general function configuration.
[SENSe:]CONTinuity:THReshold {
<
Threshold
>
|MIN|MAX|DEF}
Sets the threshold of the continuity measurement function. In this case, the instrument does not switch into continuity
mode. For activating the continuity measurement function use the FUNC command.
Parameters:
<Threshold>
0
Ω
to 1M
Ω
(adjustable in 1
Ω
steps)
MIN:
0
Ω
MAX: 1M
Ω
DEFault: 200
Ω
*RST: 200
Ω
[SENSe:]CONTinuity:THReshold? [{MIN|MAX|DEF}]
Returns the threshold of the continuity measurement function.
Return values:
<Threshold>
0
Ω
to 1M
Ω
MIN: 0.0000
MAX: 1.0006
DEFault: 2.0002