Firmware Version: 01.020 and later
English
S C P I P r o g r a m m e r s M a n u a l
H M S S e r i e s
1
HAMEG Instruments GmbH • Industriestraße 6 • D-63533 Mainhausen
Subject to change without notice
Tel.: +49 (0) 6182 800 0 • Fax: +49 (0) 6182 800 100 • E-Mail: [email protected]
Geschäftsführer: Dipl.-Ing. Holger Asmussen • AG Offenbach am Main HRB 41200
Firmware Version: 01.020 and later
English
S C P I P r o g r a m m e r s M a n u a l
H M S S e r i e s
HMO352x, HMO2524
HMO72x ... HMO202x
Firmware Version: 03.00 and later
English
:ACQuire
:CHAN1OFFSet
:DISPlay
:HCOPy
*ESR?
*OPC?
:LOGic:SIZE
:MEASure
:POD1
:SYSTem
*SRE?
S C P I P r o g r a m m e r s M a n u a l
H M O S e r i e s
SCPI Commands HMC8012
[SENSe:]CURRent:AC:BANDwidth {
<
Threshold
>
|MIN|MAX|DEF}
Sets the AC filter of the AC I measurement function.
Parameters:
<Threshold>
10 | 50 | 400
10:
10Hz filter (Slow)
50:
50Hz filter (Medium)
400:
400Hz filter (Fast)
MIN: Slow
MAX: Fast
DEF: Slow
*RST: 5.0001
[SENSe:]CURRent:AC:BANDwidth? [{MIN|MAX}]
Returns the AC filter state of the AC I measurement function.
Return values:
<State>
1.0001 | 5.01 | 4.0002
MIN:
1.0001 (Slow)
MAX:
4.0002 (Fast)
[SENSe:]CURRent:AC:NULL[:STATe] {ON|OFF}
Turns the null function for AC I measurements on or off. In this case, the instrument does not switch into AC I mode.
For activating the AC I measurement function use the FUNC command.
Parameters:
<State>
ON | OFF
ON:
Null function of AC I measurements is activated
OFF:
Null function of AC I measurements is disabled
*RST: OFF
[SENSe:]CURRent:AC:NULL[:STATe]?
Returns the NULL function state of the AC I measurement function.
Return values:
<State>
1 | 0
1
ON - NULL function activated
0
OFF - NULL function disabled