51
HAMEG Instruments GmbH • Industriestr. 6 • D-63533 Mainhausen • Deutschland
Subject to change without notice
Tel.: +49 (0) 6182 800 0 • Fax: +49 (0) 6182 800 100 • E-Mail: [email protected]
Geschäftsführer: Dipl.-Ing. Holger Asmussen, Dipl.-Ing. Andre Vander Stichelen • AG Offenbach am Main HRB 41200
1
1
HAMEG Instruments GmbH • Industriestraße 6 • D-63533 Mainhausen
Subject to change without notice
Tel.: +49 (0) 6182 800 0 • Fax: +49 (0) 6182 800 100 • E-Mail: [email protected]
Geschäftsführer: Dipl.-Ing. Holger Asmussen • AG Offenbach am Main HRB 41200
Firmware Version: 01.020 and later
English
S C P I P r o g r a m m e r s M a n u a l
H M S S e r i e s
1
HAMEG Instruments GmbH • Industriestraße 6 • D-63533 Mainhausen
Subject to change without notice
Tel.: +49 (0) 6182 800 0 • Fax: +49 (0) 6182 800 100 • E-Mail: [email protected]
Geschäftsführer: Dipl.-Ing. Holger Asmussen • AG Offenbach am Main HRB 41200
Firmware Version: 01.020 and later
English
S C P I P r o g r a m m e r s M a n u a l
H M S S e r i e s
HMO352x, HMO2524
HMO72x ... HMO202x
Firmware Version: 03.00 and later
English
:ACQuire
:CHAN1OFFSet
:DISPlay
:HCOPy
*ESR?
*OPC?
:LOGic:SIZE
:MEASure
:POD1
:SYSTem
*SRE?
S C P I P r o g r a m m e r s M a n u a l
H M O S e r i e s
SCPI Commands HMC8012
4-wire resistance configuration commands
CONFigure:FRESistance [{<Range>|AUTO|MIN|MAX|DEF}] ..................................... 51
[SENSe:]FRESistance:NULL[:STATe] {ON|OFF} .......................................................... 51
[SENSe:]FRESistance:NULL[:STATe]? ........................................................................ 51
[SENSe:]FRESistance:NULL:VALue {<Value>|MIN|MAX} .......................................... 52
[SENSe:]FRESistance:NULL:VALue? [{MIN|MAX}] .................................................... 52
[SENSe:]FRESistance:RANGe:AUTO <Mode> .......................................................... 52
[SENSe:]FRESistance:RANGe:AUTO? ....................................................................... 52
[SENSe:]FRESistance:RANGe[:UPPer] {<Range>|MIN|MAX|DEF} ............................ 52
[SENSe:]FRESistance:RANGe[:UPPer]? [{MIN|MAX}] ................................................ 53
CONFigure:FRESistance [{<Range>|AUTO|MIN|MAX|DEF}]
Configures the instrument for 4-wire resistance measurements, but does not initiate a measurement. Use the READ?
command to query the 4-wire resistance measurement value with specified range or the CONF? command to query
the 4-wire resistance function configuration.
Parameter:
<Range>
400
Ω
, 4k
Ω
, 40k
Ω
, 400k
Ω
, 4M
Ω
AUTO:
Auto range
MIN: 400
Ω
MAX: 4M
Ω
DEFault: 400
Ω
If the input signal is greater than can be measured on the selected range (manual ranging),
the instrument returns 9.9037.
*RST: AUTO
[SENSe:]FRESistance:NULL[:STATe] {ON|OFF}
Turns the null function for 4-wire (4w) resistance measurement function on or off. In this case, the instrument does not
switch into resistance mode. For activating the
Ω
measurement function use the FUNC command.
Parameters:
<State>
ON | OFF
ON:
Null function of 4w resistance measurement function is activated
OFF:
Null function of 4w resistance measurement function is disabled
*RST: OFF
[SENSe:]FRESistance:NULL[:STATe]?
Returns the null function state of the 4-wire (4w) resistance measurement function.
Return values:
<State>
1 | 0
1
ON - Null function of the 4w resistance measurement function is activated
0
OFF - Null function of the 4w resistance measurement function is disabled