89
Optional device installation
Table 12 Conductivity values at temperature for KCl solution
Solution
Temp °C
Calibration Value
to be Entered
Solution
Temp °C
Calibration Value
to be Entered
Solution
Temp °C
Calibration Value
to be Entered
30
1.099
25
1.000
20
0.902
29.8
1.095
24.8
0.996
19.8
0.898
29.6
1.091
24.6
0.992
19.6
0.894
29.4
1.087
24.4
0.988
19.4
0.890
29.2
1.083
24.2
0.984
19.2
0.886
29
1.079
24
0.980
19
0.882
28.8
1.075
23.8
0.976
18.8
0.878
28.6
1.071
23.6
0.972
18.6
0.874
28.4
1.067
23.4
0.968
18.4
0.870
28.2
1.063
23.2
0.965
18.2
0.866
28
1.059
23
0.961
18
0.862
27.8
1.055
22.8
0.957
17.8
0.858
27.6
1.051
22.6
0.953
17.6
0.854
27.4
1.047
22.4
0.949
17.4
0.850
27.2
1.043
22.2
0.945
17.2
0.846
27
1.039
22
0.941
17
0.842
26.8
1.035
21.8
0.937
16.8
0.838
26.6
1.032
21.6
0.933
16.6
0.835
26.4
1.028
21.4
0.929
16.4
0.831
26.2
1.024
21.2
0.925
16.2
0.827
26
1.020
21
0.921
16
0.823
25.8
1.016
20.8
0.917
15.8
0.819
25.6
1.012
20.6
0.913
15.6
0.815
25.4
1.008
20.4
0.909
15.4
0.811
25.2
1.004
20.2
0.905
15.2
0.807