111
Maintenance
Note: The front panel gasket has a light coating of grease to help assure a watertight seal. Do not
contaminate the grease or gasket area with dirt or debris during servicing. Always replace the gasket
if it is damaged or missing. Never reassemble the case without the gasket properly installed.
9.9 Re-installing the bottom panel
Note: Always follow the procedure below when re-installing the sampler bottom panel. Improper
panel installation may result in damage to the instrument.
1.
Hand tighten each nut until it makes contact with the panel.
2.
Tighten the nuts in sequence shown in
to 5 in.-lb (0.56 N-m).
3.
Repeat the tightening procedure in the same sequence to 10 in.-lb (1.13 N-m).
Figure 38 900MAX controller components
Item
Description
Quantity
Catalog Number
1
Fuse
1
8753
2
AA-cell battery
3
-
3
Nut
17
SE 301
4
Lock washer
17
SE 306
5
Desiccant bag
1
8849
6
O-ring
1
8606
7
Screw
17
SE 343