Basic programming setup
•
Sample distribution; deliver each sample to all bottles?—No
Bottles per sample—Yes
Number of bottles per sample—2
•
Sample volume—1000 mL
Note: A high pressure air purge is automatically applied to the intake tube between each sample
intake cycle.
Every 30 minutes a sample cycle is initiated. A 1000 mL sample is drawn into the first
bottle. the distributor arm immediately advances to bottle #2 and a 1000 mL sample is
drawn into the second bottle. Thirty minutes later, the distributor arm advances to
bottle #3, and a 1000 mL sample is taken. The distributor arm immediately advances
to bottle #4 and a 1000 mL sample is drawn into the fourth bottle. The pattern
continues each sample cycle with each set of two bottles receiving a sample each
time.
b.
Press
ACCEPT
to continue with Liquid Sensor setup.
If
YES
is selected:
Each time a sample is taken it is consecutively placed in all bottles. This allows for
split samples.
a.
Select Stop After Last Sample or Run Continuously.
b.
If Stop After Last Sample is chosen, enter the samples to collect using the
numeric keypad.
11:00 AM 21 - APR - 01
SAMPLE DISTRIBUTION
ACCEPT
RUN MODE:
CHANGE
CHOICE
RUN CONTINUOUSLY
BACKUP
CHOICES: CONTINUOUS, STOP AFTER LAST
11:00 AM 21 - APR - 01
SAMPLE DISTRIBUTION
ACCEPT
SAMPLES TO COLLECT:
BACKUP
2
CLEAR
ENTRY
ENTER: 1 — 999