Basic programming setup
13-H.
Enter the Deadband value if required or, if programming Flow Rate Of Change or
Rainfall, enter a time interval that the flow or rainfall change must take place (refer to
13-I.
Enter a delay when input is active. This delay will keep the program from starting
until the end of the delay period. Using the numeric keypad, enter the delay in
minutes and/or hours. Press
ACCEPT
.
11:00 AM 21 - APR - 01
SETPOINT SAMPLING
ACCEPT
SETPOINT SAMPLING
DEADBAND:
CANCEL
0.000
CLEAR
ENTRY
(USE NUMERIC KEYPAD)
11:00 AM 21 - APR - 01
SETPOINT SAMPLING
ACCEPT
DELAY WHEN INPUT
BECOMES ACTIVE:
CANCEL
(hrs:min)
CLEAR
ENTRY
(USE NUMERIC KEYPAD)
Table 4 Sampling triggers and settings
Channel
Sampling trigger
Settings
1
Level
High and/or Low Condition, Deadband
2
Flow
High and/or Low Condition, Deadband
3
Flow Rate of Change
High Condition within Time Interval
4
pH or ORP
High and/or Low Condition, Deadband
5
Process Temperature
High and/or Low Condition, Deadband
6
Rainfall
High Condition within Timed Interval
7
Analog Input Channel 1
High and/or Low Condition, Deadband
8
Analog Input Channel 2
High and/or Low Condition, Deadband
9
Analog Input Channel 3
High and/or Low Condition, Deadband
10
Analog Input Channel 4 or DO
High and/or Low Condition, Deadband
11
Analog Input Channel 5 or DO Temperature
High and/or Low Condition, Deadband
12
Analog Input Channel 6 or Conductivity
High and/or Low Condition, Deadband
13
Analog Input Channel 7 or Conductivity
Temperature
High and/or Low Condition, Deadband
14
External Control
Configured in External Equipment