66
Basic programming setup
the storm's early stage or “first flush.” First flush grab and main program flow-weighted
composite samples are automatically segregated. The first flush sample volume may
be set independently of the sample volume for the flow-weighted composite.
•
The special storm water program allows an external device to initiate the start of the
sampling program. A dry contact closure across Pins B and D on the Auxiliary
receptacle, held closed for at least 61 seconds, is required.
•
When sampling with multiple bottles, the sampler can be programmed to collect one
large “first flush” sample (or small multiple samples) at timed interval(s). The number
of bottles segregated for the first flush sample is selectable. Concurrently,
flow-weighted samples are collected from the beginning of the storm until all
remaining bottle(s) are filled, or after a user selected time has elapsed.
•
Storm water samplers are equipped with a special pre-rinse that occurs only with the
first sample collection. This “one time only” rinse ensures a clean intake for
installations that may remain dormant for extended periods, and extends battery life
by eliminating the pre-rinse for subsequent samples.
16-A.
Highlight Storm Water using the
UP
and
DOWN
keys on the Advanced Sampling
Menu. Press
SELECT
to continue.
16-B.
Enable or Disable Storm Water using the
CHANGE CHOICE
key. Press
ACCEPT
to
continue.
16-C.
Select a Start Condition using the
CHANGE CHOICE
key.
•
Rain
•
Level
•
Rain or Level (either condition must be met for the program to begin)
•
Rain and Level (both conditions must be met for the program
to begin)
•
Immediate (program starts as soon as the
RUN
key is pressed, a start condition is
not required)
•
External Trigger (program starts when signaled by an external device at least 61
seconds through the Auxiliary connector. No start condition is required).
16-D.
Enter the Start Condition Limits. Rainfall times will depend on the historical rainfall in
a specific area. Consult your state or an EPA regional office for details.
11:00 AM 21 - APR - 01
ADVANCED SAMPLING
SELECT
SETPOINT SAMPLING
SPECIAL PUTPUT
START/STOP TIMES
STORM WATER
RETURN
TIMED BOTTLE SETS
UPSET SAMPLE