51
Basic programming setup
a.
From the Sample Collection menu, press
CHANGE CHOICE
until
Timed
Proportional is displayed. Press
ACCEPT
to continue.
b.
Enter the Interval Between Samples. Press
ACCEPT
to continue.
c.
Select Take First Sample Immediately or After The First Interval?
Note: When the program is started, the first sample is taken immediately upon pressing the start
button or after the first interval has elapsed.
d.
Press
ACCEPT
to continue to Sample Distribution.
Flow proportional constant volume, variable time (CVVT)
CVVT sampling is a flow-proportional method of sampling. CVVT samples are taken
when a user-defined flow volume occurs in the flow stream. This occurs when the
sampler is programmed to take a sample every time a specified volume of flow occurs in
the flow stream. Specified flow volume intervals can vary with the flow rate of the stream,
therefore varied time intervals can occur when constant (fixed) volume occurs.
Flow volume is determined internally, by the optional integral flow meter or by an external
flow meter.
11:00 AM 21 - APR - 01
SAMPLE COLLECTION
ACCEPT
SAMPLE COLLECTION
CHANGE
CHOICE
TIMED PROPORTIONAL
BACKUP
CHOICES: TIMED, FLOW PROPORTIONAL
11:00 AM 21 - APR - 01
SAMPLE COLLECTION
ACCEPT
INTERVAL:
CANCEL
00:00 (hrs:min)
CLEAR
ENTRY
ENTER: 000:01—999.00 (hrs:min)
11:00 AM 21 - APR - 01
SAMPLE COLLECTION
ACCEPT
TAKE FIRST SAMPLE:
CHANGE
CHOICE
_______________
CANCEL
CHOICES: IMMEDIATE, AFTER 1
st
SAMPLE