70
Basic programming setup
18-C.
Press
CHANGE CHOICE
to select the desired channel to trigger from. Press
ACCEPT
to make the selection.
18-D.
Press either
HIGH CONDITION
or
LOW CONDITION
.
•
Flow Rate of Change and Rainfall are signals that only increase in value and do
not require Low Condition.
•
The external control signal (such as an external flow meter) must be configured in
the external equipment for the desired setpoint.
18-E.
Enter the desired high or low trigger point using the numeric keypad. Press
ACCEPT
to continue.
18-F.
Enter the Deadband value or, if programming for Flow Rate Of Change or Rainfall,
enter a time interval when the flow or rainfall change must take place (refer to
relays programming on page 100
18-G.
Enter the number of bottles (out of the total set) to set aside for upset samples.
These will be the last bottles in the tray.
18-H.
Select the method of distribution, Sample Per Bottle or
Bottles Per Sample.
11:00 AM 21 - APR - 01
UPSET SAMPLES
UPSET CONDITION:
HIGH
CONDITION
LOW
CONDITION
11:00 AM 21 - APR - 01
UPSET SAMPLES
ACCEPT
UPSET SAMPLING:
HIGH TRIGGER POINT:
CANCEL
00000 in.
CLEAR
ENTRY
(USE NUMERIC KEYPAD)
11:00 AM 21 - APR - 01
UPSET SAMPLES
ACCEPT
UPSET SAMPLING
DEADBAND:
CANCEL
0.000
CLEAR
ENTRY
(USE NUMERIC KEYPAD)
11:00 AM 21 - APR - 01
UPSET SAMPLES
ACCEPT
UPSET SAMPLING:
NUMBER OF BOTTLES
CANCEL
2
CLEAR
ENTRY
ENTER 1 — 4