Introduction
Function keys
There are three function keys that are used while operating the sampler (refer to
). These functions are dedicated keys to allow quick access. They are the white keys
located just above the numeric keypad.
3.6.2 Liquid crystal display
3.6.3 Internal humidity indicator
The internal case humidity indicator (2660) turns pink when the internal case humidity
exceeds 60%.
The sampler is equipped with an internal desiccant module (8849) to absorb any humidity
trapped in the case during final assembly. Under normal operating conditions, this
desiccant provides long-term protection against condensed moisture inside the case.
Replacement of the internal desiccant module is only necessary if the indicator turns
pink. Refer to
for details.
Table 1 Function key descriptions
Main Menu
This is the starting point to get to any other point in the program. Press the Main Menu key at any time during
programming to return to the Main Menu Screen. The current action is cancelled if changes are not yet accepted.
Manual Mode
Manually controls the operation of the sample pump and the distributor arm.
ADVANCED DISTRIBUTOR
soft key: Moves the distributor arm to the user selected bottle. Used to verify the operation of
the distributor or when repositioning the arm if it was moved by hand.
GRAB SAMPLE
soft key: Takes a sample in the same manner as when a program is running. Includes all pre-rinses and
sample retries, if programmed.
PUMP OPERATION
soft key: Allows manual control of the pump in both forward and reverse directions. Once started, the
pump is stopped by pressing any key.
Run/Stop
Runs (or resumes) a program and stops a currently running program.
Table 2 LCD displays
Menu Bar
The Menu Bar appears in a black band on the top edge of the display. The upper left corner of the menu bar
shows the time and date. The upper right corner shows the name of the current menu (
).
Status Bar
The Status Bar appears along the bottom edge of the display. The appearance of the status bar changes
depending upon the function performed (
). The lower left corner of the Status Bar indicates whether a
program is Complete, Running, Halted, or Ready To Start. If it is not needed during a programming step, it
disappears. The lower right corner displays system alarm conditions, such as low memory battery, jammed
distributor etc. For a list of possible alarms refer to
. The status bar also lists the valid
choices when entering certain programming information.