Total Oxidants
50081_TotalOxidants.fm
Page 119
Total Oxidants
*
• Soak all glassware in a strong chlorine dioxide (ClO
2
) solution (300–500 mg/L) for at least 1 hour.
Rinse thoroughly with deionized water. Use the glassware for this method only.
• Chlorine dioxide and its by-products are volatile and can be easily lost from aqueous solution. Minimize air contact by attaching a
flexible hose to a tap and placing the end at the bottom of a 1-L amber glass bottle. Turn on the tap and allow several volumes to
overflow, then slowly remove the sample line and cap the container with minimum headspace.
• Minimize agitation when measuring sample volumes. Remove sample portions with a volumetric pipet. Always place the tip at the
bottom of the sample container. If using 200-mL sample increments, use a 100-mL pipet to withdraw two portions of sample.
• Always use organic-free water for sample dilution (Cat. No. 26415-49).
• Rinse the electrode, temperature probe, and anti-diffusion tip with deionized water before every titration.
.
• Download the “Certificate of Analysis” (COA) to obtain the exact concentration of any unopened bottle of Hach titrant or standard
solution. See
Obtaining the Certificate of Analysis for Hach Reagents
.
• Hach buffer reagents for chlorine titrations are highly recommended for this analysis.
• Never substitute buffers designed for calibrating pH meters. They may contain dyes that interfere with amperometric titration.
• Never use buffers contaminated with mold or bacteria.
• Adjust the contrast of the display by accessing the Methods menu and pressing
7
to lighten or
0
to darken.
• A distinction is drawn between running a new Test and a new Sample. Each Test is a replicate run of the current or previous
sample analysis. The AutoCAT 9000 automatically tracks the results of a series of tests, and automatically calculates the mean
and standard deviation for all the results. Each Test is given a different number, all of which are listed under the same Sample ID.
When starting a new Sample, a new ID is assigned.
• The AutoCAT 9000 can accommodate any sample size.
Standard Methods
recommends a default volume of 200 mL. The
volume can be adjusted to expand the test range.
• The AutoCAT 9000 calculates concentration based on the sample volume. Make sure that the sample volume is correct.
• Press
STOP
at any time to interrupt instrument operation; the user can resume, restart, or quit completely.
• Press the
CHECK MARK
key at any time during data acquisition to halt data collection.
• Press
DEL
during any timer period (except Start Timer) to immediately exit the timer and proceed to the next step.
• Press
ESC
during any titration/calibration data collection period or timer to access Edition mode. See
.
• Flush the burette each day before the first sample test or calibration is performed. See
• Flush the burette when changing titrants. See
* Adapted from Gates, “The Chlorine Dioxide Handbook”, AWWA, 1998. Not suitable for compliance testing.
Ch Chlorine Dioxide
“Quick” Two Step Amperometric Titration
Titration Workstation
0.1–5.0 mg/L as Cl0
2
+ ClO
2
–
Scope and Application:
For drinking water
Sample Tips and Techniques
Reagent Tips and Techniques
Instrument Tips and Techniques
Summary of Contents for AutoCAT 9000
Page 2: ...50081_TTL fm...
Page 8: ...Page 6 50081_Safety fm...
Page 14: ...Page 12 50081_Introduction fm...
Page 24: ...Page 22 50081_OperationStop fm...
Page 34: ...Basic Operation Basic Operation Page 32 50081_Operation fm...
Page 44: ...Page 42 50081_SoftwareSetup fm...
Page 51: ...Calibration Procedures 50081_CalibrationStopper fm Page 49 Section 6 Calibration Procedures...
Page 52: ...Page 50 50081_CalibrationStopper fm...
Page 66: ...Page 64 50081_Sampling fm...
Page 67: ...Titration Procedures 50081_ProcedureStopper fm Page 65 Section 8 Titration Procedures...
Page 68: ...Page 66 50081_ProcedureStopper fm...
Page 138: ...Page 136 50081_ManualEndpoint fm...
Page 140: ...Page 138 50081_BDL fm...
Page 154: ...Page 152 50081_Titration fm...
Page 158: ...Page 156 50081_PrinterInstallation fm...
Page 160: ...Page 158 50081_ReplacementParts fm...
Page 170: ...Page 168 50081_GeneralInfoStop fm...
Page 176: ......