![Hach AutoCAT 9000 Instruction Manual Download Page 71](http://html1.mh-extra.com/html/hach/autocat-9000/autocat-9000_instruction-manual_3154724071.webp)
Chlorine Dioxide Generator Yield
Chlorine Dioxide Generator Yield
50081_ChlorineDioxideGeneratorYield.fm
Page 69
9.
Highlight Predilution
volume.
Press the
CHECK MARK
and
enter the volume pipetted in
Step 4.
Press
1
to confirm.
Note:
Final dilution volume and
Volume for Test default values
need not be changed.
10.
Add 1 dose of pH 7.0
Phosphate Buffer.
Press the
CHECK MARK
to confirm.
Note:
Phosphate buffer is
added to adjust the sample pH.
The precise amount added is
not crucial for the accuracy or
precision of the analysis.
11.
Use the measuring
spoon to add 1.0 g of
potassium iodide (KI).
Press the
CHECK MARK
to confirm.
Note:
KI is added in excess.
The precise amount added is
not crucial for the accuracy or
precision of the analysis.
12.
The timer will start,
allowing the reagents to mix
before starting the titration.
Summary of Contents for AutoCAT 9000
Page 2: ...50081_TTL fm...
Page 8: ...Page 6 50081_Safety fm...
Page 14: ...Page 12 50081_Introduction fm...
Page 24: ...Page 22 50081_OperationStop fm...
Page 34: ...Basic Operation Basic Operation Page 32 50081_Operation fm...
Page 44: ...Page 42 50081_SoftwareSetup fm...
Page 51: ...Calibration Procedures 50081_CalibrationStopper fm Page 49 Section 6 Calibration Procedures...
Page 52: ...Page 50 50081_CalibrationStopper fm...
Page 66: ...Page 64 50081_Sampling fm...
Page 67: ...Titration Procedures 50081_ProcedureStopper fm Page 65 Section 8 Titration Procedures...
Page 68: ...Page 66 50081_ProcedureStopper fm...
Page 138: ...Page 136 50081_ManualEndpoint fm...
Page 140: ...Page 138 50081_BDL fm...
Page 154: ...Page 152 50081_Titration fm...
Page 158: ...Page 156 50081_PrinterInstallation fm...
Page 160: ...Page 158 50081_ReplacementParts fm...
Page 170: ...Page 168 50081_GeneralInfoStop fm...
Page 176: ......