Chlorine Dioxide
Chlorine Dioxide
50081_Chlorine_ChlorineDioxide.fm
Page 83
9.
Add 1.0 mL of pH 7.0
Phosphate Buffer Solution.
Press the
CHECK MARK
to confirm.
Note:
Phosphate buffer is
added to adjust the sample pH.
The precise amount added is
not crucial for the accuracy or
precision of the analysis.
10.
Use the measuring
spoon to add 1.0 g of
potassium iodide (KI).
Press the
CHECK MARK
to confirm.
Note:
KI is added in excess.
The precise amount added is
not crucial for the accuracy or
precision of the analysis.
11.
The timer will start.
This allows a short period of
time for the reagents to mix
before starting the titration.
12.
Lower the electrode
head onto the beaker rim.
Press the
CHECK MARK
to confirm.
13.
The analysis parameters
that have been stored for the
current Sub-ID may be
edited at this time.
Highlight the appropriate
line and press the
CHECK MARK
to change one
of the analysis parameters or
press
1
to confirm.
Note:
Any changes made will
become the new default
parameters for this method
Sub-ID.
Note:
These Sub-ID
parameters can be edited here
as 'run-time' settings, or from
the Main Menu as described in
Section 3.5.
Note:
This step will be skipped
if auto-scaling software is
enabled.
14.
The Start Timer
will start to allow the
electrode to stabilize before
data collection.
Once the electrode has
stabilized, the AutoCAT will
begin data acquisition and
plot the titration curve.
15.
The Zoom Range
Determination window will
display the titration curve.
Use the navigation buttons to
select and move the cursors
to define the range of data
points for analysis.
Note:
Defining the range of
data points excludes
undesirable portions of the
titration curve from
the analysis.
16.
Select type of end point
determination.
Press
DEL
for Manual.
Note:
Automatic determination
attempts to identify linear
segments without user input
and is
not
recommended
for
this method.
Summary of Contents for AutoCAT 9000
Page 2: ...50081_TTL fm...
Page 8: ...Page 6 50081_Safety fm...
Page 14: ...Page 12 50081_Introduction fm...
Page 24: ...Page 22 50081_OperationStop fm...
Page 34: ...Basic Operation Basic Operation Page 32 50081_Operation fm...
Page 44: ...Page 42 50081_SoftwareSetup fm...
Page 51: ...Calibration Procedures 50081_CalibrationStopper fm Page 49 Section 6 Calibration Procedures...
Page 52: ...Page 50 50081_CalibrationStopper fm...
Page 66: ...Page 64 50081_Sampling fm...
Page 67: ...Titration Procedures 50081_ProcedureStopper fm Page 65 Section 8 Titration Procedures...
Page 68: ...Page 66 50081_ProcedureStopper fm...
Page 138: ...Page 136 50081_ManualEndpoint fm...
Page 140: ...Page 138 50081_BDL fm...
Page 154: ...Page 152 50081_Titration fm...
Page 158: ...Page 156 50081_PrinterInstallation fm...
Page 160: ...Page 158 50081_ReplacementParts fm...
Page 170: ...Page 168 50081_GeneralInfoStop fm...
Page 176: ......