Page 16
50081_Assembly.fm
Operating Environment
2.2.3 Connecting the Titrant Bottle
2.
Clip the stop ring into the locating hole on the support rod.
Note:
The stop ring may be pre-assembled on the support rod.
3.
Compress the clips on each side of the electrode head and slide the
head onto the metal rod. Release the two clips to secure the head
in place.
4.
Drop the supplied magnetic stir bar into the titration beaker and place
the beaker on the titration platform.
Note:
The provided stir bar has been specifically selected for use with the
AutoCAT 9000. A different stir bar may spin erratically and produce
unwanted turbulence and added noise in the titration curve.
1.
Place the titrant bottle into the bottle holder.
2.
Thread the desiccant column onto the supplied bottle stopper.
3.
Thread the appropriate bottle stopper (for 500 mL or 1000 mL titrant
bottle) onto the new titrant bottle and place it into the holder.
Summary of Contents for AutoCAT 9000
Page 2: ...50081_TTL fm...
Page 8: ...Page 6 50081_Safety fm...
Page 14: ...Page 12 50081_Introduction fm...
Page 24: ...Page 22 50081_OperationStop fm...
Page 34: ...Basic Operation Basic Operation Page 32 50081_Operation fm...
Page 44: ...Page 42 50081_SoftwareSetup fm...
Page 51: ...Calibration Procedures 50081_CalibrationStopper fm Page 49 Section 6 Calibration Procedures...
Page 52: ...Page 50 50081_CalibrationStopper fm...
Page 66: ...Page 64 50081_Sampling fm...
Page 67: ...Titration Procedures 50081_ProcedureStopper fm Page 65 Section 8 Titration Procedures...
Page 68: ...Page 66 50081_ProcedureStopper fm...
Page 138: ...Page 136 50081_ManualEndpoint fm...
Page 140: ...Page 138 50081_BDL fm...
Page 154: ...Page 152 50081_Titration fm...
Page 158: ...Page 156 50081_PrinterInstallation fm...
Page 160: ...Page 158 50081_ReplacementParts fm...
Page 170: ...Page 168 50081_GeneralInfoStop fm...
Page 176: ......