Software Setup
Software Setup
Page 38
50081_SoftwareSetup.fm
1.
From the Titrants
menu,
press
2
to select Install
Titrants.
2.
Press
1
to select
Titrants Setup.
3.
Highlight Lot no.
Press the
CHECK MARK
to
input or change.
4.
Use the alphanumeric
keypad to input or change
the titrant lot number—
located next to the barcode
on the titrant bottle.
Press the
CHECK MARK
to
validate.
5.
Highlight Expiration.
Press the
CHECK MARK
to
input or change.
6.
Use the alphanumeric
keypad to input the month
and year of the titrant
expiration date (located next
to the barcode on the titrant
bottle).
Press the
CHECK MARK
to
validate.
7.
Verify the titrant
information. Press
1
to
confirm.
8.
Press
2
to select
No Rinse.
Note:
Initial installation does
not require a burette rinse;
however, it is highly
recommended for titrant
replacement or exchange.
4.4 Software Setup for Titrant Installation
Summary of Contents for AutoCAT 9000
Page 2: ...50081_TTL fm...
Page 8: ...Page 6 50081_Safety fm...
Page 14: ...Page 12 50081_Introduction fm...
Page 24: ...Page 22 50081_OperationStop fm...
Page 34: ...Basic Operation Basic Operation Page 32 50081_Operation fm...
Page 44: ...Page 42 50081_SoftwareSetup fm...
Page 51: ...Calibration Procedures 50081_CalibrationStopper fm Page 49 Section 6 Calibration Procedures...
Page 52: ...Page 50 50081_CalibrationStopper fm...
Page 66: ...Page 64 50081_Sampling fm...
Page 67: ...Titration Procedures 50081_ProcedureStopper fm Page 65 Section 8 Titration Procedures...
Page 68: ...Page 66 50081_ProcedureStopper fm...
Page 138: ...Page 136 50081_ManualEndpoint fm...
Page 140: ...Page 138 50081_BDL fm...
Page 154: ...Page 152 50081_Titration fm...
Page 158: ...Page 156 50081_PrinterInstallation fm...
Page 160: ...Page 158 50081_ReplacementParts fm...
Page 170: ...Page 168 50081_GeneralInfoStop fm...
Page 176: ......